Measuring LED light - do I need a special light meter?

All of Extech's light meters can work with LED lights when they are used to compare different light sources. Let me explain:Our light meters measure visible light in the Human Photopic range.That…Continue

Started Dec 11, 2013

RHT10 - Logger Setup will not remember my previous settings

When a user opens the RHT10 software and runs Logger Setup, the values they enter should be remembered for the next time.  This makes it easier for multiple loggers to be setup in the same way…Continue

Started Dec 4, 2013

TH10 - Will not connect

A few people have reported that when they received their TH10 temperature datalogger and proceeded to load the software, they found that either they were unable to load the software or when they did…Continue

Tags: TH10

Started Mar 22, 2013

Gifts Received

Gift

Steve Hyde has not received any gifts yet

Give a Gift

 

Welcome, Steve Hyde!

Latest Activity

Steve Hyde's discussion was featured

CT70/CT80 How does the Voltagge Drop test work without tripping the circuit breaker?

The CT tester checks the voltage drop on the power line by applying a short circuit to the power line at a low voltage point on the rising voltage cycle.  The meter measures the Peak current drawn by that short circuit test.  This measured peak current is compared to the Expected max current of the circuit.  The resulting percentage…See More
Feb 14, 2014
Steve Hyde posted a discussion

CT70/CT80 How does the Voltagge Drop test work without tripping the circuit breaker?

The CT tester checks the voltage drop on the power line by applying a short circuit to the power line at a low voltage point on the rising voltage cycle.  The meter measures the Peak current drawn by that short circuit test.  This measured peak current is compared to the Expected max current of the circuit.  The resulting percentage…See More
Feb 14, 2014
Steve Hyde posted a discussion

Measuring LED light - do I need a special light meter?

All of Extech's light meters can work with LED lights when they are used to compare different light sources. Let me explain:Our light meters measure visible light in the Human Photopic range.That means they measure light that the human eye can detect. LEDs also produce a "spike" of light in  the 410nm range (blue light) which is just outside of the Human eye detection area. If you are comparing light sources, Fluorescent to incandescent, to Sodium, to LED, etc. any of our light meters will…See More
Dec 11, 2013
Steve Hyde posted a discussion

RHT10 - Logger Setup will not remember my previous settings

When a user opens the RHT10 software and runs Logger Setup, the values they enter should be remembered for the next time.  This makes it easier for multiple loggers to be setup in the same way without having to perform multiple setup actions over again. When the RHT10 software loads, it places a file called  loggerconfig.ini  into the C:\WINDOWS folder. When a user edits the RHT10 logger setup, this "ini" file gets updated with the newest settings or I say it "should" get updated.Some recent…See More
Dec 4, 2013
Steve Hyde posted a discussion

TH10 - Will not connect

A few people have reported that when they received their TH10 temperature datalogger and proceeded to load the software, they found that either they were unable to load the software or when they did the TH10 would not connect to their PC. A number of CDs that went out with the TH10 dataloggers were found corrupted or missing the driver. The solution is to go to the Software downloads page on the Extech web site and download the TH10 Version 5.0 software. SAVE it to your hard drive, Unzip it…See More
Mar 22, 2013
Steve Hyde posted a discussion

RH520A temperature and humidity datalogger - wrong dates

We occasionally get a call on the RH520A, that the data has the wrong time and dates (2004).Why does this occur? When the RH520A is first powered on, by either batteries or AC power, the CPU immediately begins to record temperature and humidity data at the default CPU start date/time of 1/1/2004 00:00.  The data is recorded off of this initial date/time and sample rate (1.0 minute). To get your meter to record in the present;:After power on, set your Date, Time, and Sample rate, and then press…See More
Mar 12, 2013
Steve Hyde's discussion was featured

How do I convert my Refractometer reading in Brix to a percentage

This is a tough one as it makes a difference depending on the type of material (Dissolved solid) in solution.Brix can be converted to Refractive Index, but to calculate a % of some material, there are other factors to consider. For a basic sugar in water solution it is simple:Brix = the mass of sugar relative to the mass of the sample.22 Brix = 22 grams of sugar in  78 grams of waterTotal mass = 100 gramsDivide 100 grams into the mass of the solid (Sugar) and you get 0.22 or 22%With a straight…See More
Feb 27, 2013
Steve Hyde posted a discussion

How do I convert my Refractometer reading in Brix to a percentage

This is a tough one as it makes a difference depending on the type of material (Dissolved solid) in solution.Brix can be converted to Refractive Index, but to calculate a % of some material, there are other factors to consider. For a basic sugar in water solution it is simple:Brix = the mass of sugar relative to the mass of the sample.22 Brix = 22 grams of sugar in  78 grams of waterTotal mass = 100 gramsDivide 100 grams into the mass of the solid (Sugar) and you get 0.22 or 22%With a straight…See More
Nov 27, 2012
Steve Hyde's discussion was featured

Panel Mount Techometer - Proximity sensor

Here are the mounting dimensions for the metal proximity sensor (p/n 461955) available for the Panel mount Tachometer (461950)  - attached. These dimensions are not posted in the manual.461955%20%20Proximity%20sensor%20-%20dimensions.docSee More
Oct 25, 2012
Steve Hyde posted a discussion

Panel Mount Techometer - Proximity sensor

Here are the mounting dimensions for the metal proximity sensor (p/n 461955) available for the Panel mount Tachometer (461950)  - attached. These dimensions are not posted in the manual.461955%20%20Proximity%20sensor%20-%20dimensions.docSee More
Oct 25, 2012
Steve Hyde updated their profile
Oct 25, 2012
Steve Hyde's discussion was featured

42260, 42270, 42280 and Windows 7

Currently, the driver for the 42280 Temperature datalogger is not up to date for Windows 7.  Here is how to get these Temperature and humidity Datalogger running on Windows 7.Make sure you are running Version 3.5 software for the 42280.If not, Download the version 3.5 42280 Software (TrLog) from the Extech web site and install it. Download the Prolific PL-2303 driver from the ProlificUSA web site. (…See More
Aug 2, 2012
Steve Hyde posted a discussion

42260, 42270, 42280 and Windows 7

Currently, the driver for the 42280 Temperature datalogger is not up to date for Windows 7.  Here is how to get these Temperature and humidity Datalogger running on Windows 7.Make sure you are running Version 3.5 software for the 42280.If not, Download the version 3.5 42280 Software (TrLog) from the Extech web site and install it. Download the Prolific PL-2303 driver from the ProlificUSA web site. (…See More
Aug 1, 2012
Steve Hyde's discussion was featured

407764 - Sound Level Meter - New Version of Software available

407764 Sound meter – New Version 7.0 Software ReleaseThis software will replace version 6.2 and is now compatible with Vista, XP, Windows 7 32-bit and Windows 7 64-bit PCs.You can download a copy from the Extech Software Downloads web page.http://www.extech.com/instruments/software.aspNote: a Run-time error was noted by one customer when he tried to download recorded data and the date and time on his meter did not match his PC.This is…See More
Jul 12, 2012
Steve Hyde posted a discussion

407764 - Sound Level Meter - New Version of Software available

407764 Sound meter – New Version 7.0 Software ReleaseThis software will replace version 6.2 and is now compatible with Vista, XP, Windows 7 32-bit and Windows 7 64-bit PCs.You can download a copy from the Extech Software Downloads web page.http://www.extech.com/instruments/software.aspNote: a Run-time error was noted by one customer when he tried to download recorded data and the date and time on his meter did not match his PC.This is…See More
Jul 12, 2012
Steve Hyde is now a member of Planet Extech
May 10, 2012

Comment Wall

You need to be a member of Planet Extech to add comments!

Join Planet Extech

  • No comments yet!
 
 
 

Photos

Loading…
  • Add Photos
  • View All

Members

Latest Activity

© 2017   Created by Vatche Arabian.   Powered by

Badges  |  Report an Issue  |  Terms of Service